TT-F1800 Full-automatic Total Migration Tester Feature:
Feature of independent cooling and heating box system
- Independent cooling and heating box, one box for heating, another box for cooling, save cold and heat exchange time
- The unique design of automatic temperature control and automatic exhaust of moisture.
- It can control the temperature and humidity when weighing, which ensure the consistency of ambient temperature and humidity during each weighing.
- Temperature control range: temperature control range of water bath: 68.3 ~ 100 ℃; temperature control range of heating box: 68.3 ~ 117.9 ℃; Cooling box temperature control range: < 27.5 ℃; Accurate to ±3 ℃
Features of water bath built-in system
- Built-in water bath evaporation system, add water to water bath automatically and auto detection for water level, auto control for water bath temperature, auto evaporate and drying;
- Over-temperature automatic alarm system and water-dry automatic alarm system, multiple safety design, full protection of safe treatment of water temperature, water bath water, solvents emissions;
Solvents recovery system
- Solvent evaporates in a fully enclosed system, automatic recovery of exhaust gas, no leakage
- Configure the solvent recovery system, the chamber adopts fully enclosed design, effectively prevent solvent leakage, which ensure the health of operators.
Nitrogen-filled protection
- The system is equipped with carrier gas to support rapid evaporation testing.
Negative Pressure Protection for testing chamber
- Entrance and exit of the testing chamber are totally-enclosed
- The test chamber can vacuum at any time to guarantee negative pressure.
Full-automatic Test
- Integrated testing process with water bath, drying, weighing and evaporation all-in-one instrument, with built-in operation system, which carry ARM control system to control the mainframe by touch screen (testing equipment and control software combined into one).
- Unattended automatic constant weight, reservation constant weight of empty cup, effectively shorten test time and improve test efficiency.
- Fully automatic test, the test process can be set flexibly, one button operation, automatic feeding, automatic evaporation constant weight, automatic judgment, and stop.
- With 18 test stations can realize vaporization independently, can test different samples at the same time and comes out independent data.
Software system
- The software is designed according to the requirements of the new GMP appendix computer system.
- User name and password are required to login workstation.
- Users are divided into various levels such as operators and administrators (such as but not limited to administrators, operators, observers, etc.)
- Administrators can adjust permissions of various levels; for example, increase or decrease system control of one level.
- With audit trail function (system audit trail, project operation audit trail and method audit trail), each data changes will be recorded; the security and integrity of test data are guaranteed.
- Real-time display test weight, migration, temperature and other data of each chamber.
- Test report — Display detailed test data and curves. (Flexibility to choose which curves and parts of a curve to report). Reports can be exported in Office, PDF and other formats.
Multiple security considerations
- Water heating instead of electric heating, heating system is set out of testing chamber, no high voltage fire in evaporation chamber, security is guaranteed.
- Over-temperature alarm, evaporation complete alarm, multiple alarm setting, good security performance.
TT-F1800 Full-automatic Total Migration Tester Technical Parameters:
Items | Technical Parameters |
Test range | 0.2mg/dm2-167000mg/dm2 |
Resolution | 0.0001 g/dm2 |
Scale technical parameters | measurement range:0~200g;resolution:0.1mg |
Temperature control range of water bath | Room temperature~100℃ |
Temperature control range of heating box | Room temperature~120℃ |
Temperature control range of cooling box | Room temperature |
Temperature accuracy | ±0.5℃ |
Chamber temperature uniformity | ±5℃ |
Quantity of samples | 1-18pcs(with independent data) |
Constant weight time | 2.5h(conventional) |
Evaporating dish volume | 0-200mL |
Dimensions | L×W×H:180cm×90cm×120cm |
Power supply | 220V,50Hz |
Power | 5000W |
TT-F1800 Full-automatic Total Migration Tester Configuration List:
A, Spare parts | ||||
NO. | Name | Remark | Quantity | |
1 | Balance module | 0~220g | 2 PCS | |
2 | Rotary tray | 19 holes | 1 PC | |
3 | Evaporation beaker | Φ85mm | 40 PCS | |
4 | Funnel | Φ85mm | 40 PCS | |
5 | Cup cover | Auxiliary funnel | 20 PCS | |
6 | Standard weight | 200g | 1 PC | |
7 | Power cable | 1 PC | ||
8 | Recovery system control line | 1 PC | ||
9 | Air switch | 1 PC | ||
10 | Gas pipe | Φ8 | 1 PC | |
11 | pressure reducing valve | 1 PC | ||
12 | Cleaning spray | 400ml | 1 PC | |
B, Optional parts | ||||
NO. | Name | Unit | Quantity | Remark |
1 | Nitrogen generator | Optional | ||
2 | Air compressor | Optional |